4-50
Programming Commands
CALCulate3 Subsystem
DELTa:POWer[:STATe]
Turns the delta- power measurement mode on and off.
Syntax
:CALCulate3:DELTa:POWer[:STATe]{?| {ON | OFF | 1 | 0}}
Attribute
Summary
Preset
State: off
*RST State: off
SCPI Comp liance: instrument sp ecific
Description
When this state is on, the p ower of the reference laser line is sub-
tracted from the p ower values of all laser lines excep t the reference.
The p ower data returned by the CALC3:DATA? query is the array of
laser line power levels normalized to the power level of the reference
laser line. The power of the reference laser line is returned as an
absolute p ower (unnormalized). The frequency data returned is the
array of absolute frequency values.
DELTa:PRESet
Turns off all delta measurement states.
Syntax
:CALCulate3:DELTa:PRESet
Attribute
Summary
Preset
State: not affected
*RST State: not affected
SCPI Comp liance: instrument sp ecific
Command Only
Note
Only one STATe command can be turned on at any one time. Attempting to turn
more than one state on at a time results in a “–221 Settings Conflict” error.
Refer
to “Measure delta, drift, and signal-to-noise” on page 3-14
for additional informa-
tion on selecting measurements.
Summary of Contents for 86120C
Page 12: ......
Page 26: ...1 14 Getting Started Returning the Instrument for Service ...
Page 27: ...1 15 Getting Started Returning the Instrument for Service ...
Page 28: ......
Page 96: ...3 18 Programming Monitoring the Instrument ...
Page 128: ...3 50 Programming Lists of Commands ...
Page 236: ...4 108 Programming Commands UNIT Subsystem ...
Page 248: ......
Page 264: ...6 16 Specifications and Regulatory Information Product Overview ...
Page 269: ...7 5 Reference Menu Maps Appl s Menu ...
Page 271: ...7 7 Reference Menu Maps Display List by WL Menu Delta On Menu ...
Page 273: ...7 9 Reference Menu Maps System Print Menu ...
Page 274: ...7 10 Reference Menu Maps System Setup Menu ...
Page 284: ......
Page 292: ......
Page 293: ......