4-53
Programming Commands
CALCulate3 Subsystem
DELTa:REFerence:WNUMber
Selects the reference laser line for delta calculations.
Syntax
:CALCulate3:DELTa:REFerence:WNUMber{?| {<real> | MINimum | MAXimum}}
<real>
is a wave number value that is within the following limits:
Constant
Description
MINimum 6,060
cm
-1
(1650 nm)
MAXimum 7,824
cm
-1
(1270 nm)
Attribute
Summary
Preset
State: 7,824 cm
- 1
(1270 nm)
*RST State: 7,824 cm
- 1
(1270 nm)
SCPI Comp liance: instrument sp ecific
Description
The reference will be the laser line at the wave number closest to the
wave number entered. Subsequent measurements will use the wave
number closest to the reference wave number used for the p revious
measurement.
The query returns the current wave number of the reference laser line.
The default units for the
<real>
p arameter are m
–1
.
Summary of Contents for 86120C
Page 12: ......
Page 26: ...1 14 Getting Started Returning the Instrument for Service ...
Page 27: ...1 15 Getting Started Returning the Instrument for Service ...
Page 28: ......
Page 96: ...3 18 Programming Monitoring the Instrument ...
Page 128: ...3 50 Programming Lists of Commands ...
Page 236: ...4 108 Programming Commands UNIT Subsystem ...
Page 248: ......
Page 264: ...6 16 Specifications and Regulatory Information Product Overview ...
Page 269: ...7 5 Reference Menu Maps Appl s Menu ...
Page 271: ...7 7 Reference Menu Maps Display List by WL Menu Delta On Menu ...
Page 273: ...7 9 Reference Menu Maps System Print Menu ...
Page 274: ...7 10 Reference Menu Maps System Setup Menu ...
Page 284: ......
Page 292: ......
Page 293: ......