4-76
Programming Commands
DISPlay Subsystem
MARKer:MAXimum
Sets the marker to the laser line that has the maximum p ower.
Syntax
:DISPlay:MARKer:MAXimum
Attribute
Summary
Preset
State: marker set to maximum- p ower laser line
*RST State: marker set to maximum- p ower laser line
SCPI Comp liance: instrument sp ecific
Command Only
MARKer:MAXimum:LEFT
Moves the marker left to the next laser line.
Syntax
:DISPlay:MARKer:MAXimum:LEFT
Attribute
Summary
Preset
State: marker set to maximum- p ower laser line
*RST State: marker set to maximum- p ower laser line
SCPI Comp liance: instrument sp ecific
Command Only
Description
Moves the marker from the current marker p osition to the next laser
line having the following characteristic:
• shorter wavelength
• lower frequency
• lower wave number
If the display is in the
List by Ampl
mode, it will be changed to
List by
WL
before the marker is moved.
Summary of Contents for 86120C
Page 12: ......
Page 26: ...1 14 Getting Started Returning the Instrument for Service ...
Page 27: ...1 15 Getting Started Returning the Instrument for Service ...
Page 28: ......
Page 96: ...3 18 Programming Monitoring the Instrument ...
Page 128: ...3 50 Programming Lists of Commands ...
Page 236: ...4 108 Programming Commands UNIT Subsystem ...
Page 248: ......
Page 264: ...6 16 Specifications and Regulatory Information Product Overview ...
Page 269: ...7 5 Reference Menu Maps Appl s Menu ...
Page 271: ...7 7 Reference Menu Maps Display List by WL Menu Delta On Menu ...
Page 273: ...7 9 Reference Menu Maps System Print Menu ...
Page 274: ...7 10 Reference Menu Maps System Setup Menu ...
Page 284: ......
Page 292: ......
Page 293: ......