4-85
Programming Commands
SENSe Subsystem
CORRection:ELEVation
Sets the elevation value used by the instrument to comp ensate for air
dispersion.
Syntax
:SENSe:CORRection:ELEVation{?| {<integer> | MINimum | MAXimum}}
<integer>
is the altitude in meters.
Constant
Description
MINimum 0
m
MAXimum 5000
m
Attribute
Summary
Non- sequential command
Preset
State: unaffected by
*RST sets this value to the minimum.
SCPI Comp liance: instrument sp ecific
0Description
Changing the elevation value causes the current data to be rep ro-
cessed.
Query Response
The query form returns the current elevation setting as shown in the
following example:
+1500
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command has
the time to complete before sending any more commands to the instrument.
Refer
to “Always force the Agilent 86120C to wait for non-sequential commands” on
page 3-12
for more information.
Summary of Contents for 86120C
Page 12: ......
Page 26: ...1 14 Getting Started Returning the Instrument for Service ...
Page 27: ...1 15 Getting Started Returning the Instrument for Service ...
Page 28: ......
Page 96: ...3 18 Programming Monitoring the Instrument ...
Page 128: ...3 50 Programming Lists of Commands ...
Page 236: ...4 108 Programming Commands UNIT Subsystem ...
Page 248: ......
Page 264: ...6 16 Specifications and Regulatory Information Product Overview ...
Page 269: ...7 5 Reference Menu Maps Appl s Menu ...
Page 271: ...7 7 Reference Menu Maps Display List by WL Menu Delta On Menu ...
Page 273: ...7 9 Reference Menu Maps System Print Menu ...
Page 274: ...7 10 Reference Menu Maps System Setup Menu ...
Page 284: ......
Page 292: ......
Page 293: ......