4-86
Programming Commands
SENSe Subsystem
CORRection:MEDium
Sets the Agilent 86120C to return wavelength readings in a vacuum or
standard air.
Syntax
:SENSe:CORRection:MEDium{?| {AIR | VACuum}}
Argument
Description
AIR
Selects wavelength values in standard air.
VACuum
Selects wavelength values in a vacuum.
Attribute
Summary
Preset
State: VAC
*RST State: VAC
SCPI Comp liance: instrument sp ecific
Description
Standard air is defined to have the following characteristics:
Barometric pressure: 760 torr
Temperature:
15
°
C
Relative humidity:
0%