4-100
Programming Commands
SYSTem Subsystem
PRESet
Performs the equivalent of p ressing the front- p anel
PRESET
key.
Syntax
:SYSTem:PRESet
Attribute
Summary
Preset
State: none
*RST State: none
SCPI Compliance: standard
Command Only
Description
The instrument state is set according to the settings shown in the fol-
lowing table.
Table 4-19. Instrument Conditions (1 of 2)
Item
Settings after Preset
Key Pressed
Settings after Power
Turned On
Display mode
single wavelength
last state
a
Wavelength range limiting
on
last state
Start wavelength
1270 nm
last state
Stop wavelength
1650 nm
last state
Graphical display
off
off
Measurement acquisition
continuous
last state
Wavelength calibration
vacuum
last state
Elevation correction value
not affected
last state
Wavelength units
nm
last state
Amplitude units
dBm
last state
Power offset
0 dB
last state
Peak threshold
10 dB
last state
Peak excursion
15 dB
last state
Summary of Contents for 86120C
Page 12: ......
Page 26: ...1 14 Getting Started Returning the Instrument for Service ...
Page 27: ...1 15 Getting Started Returning the Instrument for Service ...
Page 28: ......
Page 96: ...3 18 Programming Monitoring the Instrument ...
Page 128: ...3 50 Programming Lists of Commands ...
Page 236: ...4 108 Programming Commands UNIT Subsystem ...
Page 248: ......
Page 264: ...6 16 Specifications and Regulatory Information Product Overview ...
Page 269: ...7 5 Reference Menu Maps Appl s Menu ...
Page 271: ...7 7 Reference Menu Maps Display List by WL Menu Delta On Menu ...
Page 273: ...7 9 Reference Menu Maps System Print Menu ...
Page 274: ...7 10 Reference Menu Maps System Setup Menu ...
Page 284: ......
Page 292: ......
Page 293: ......