Common Commands
5
-
9
In this example, the command enables ESB (event sum-
mary) bit 5 in the status byte register to generate a ser-
vice request.
*STB?
The *STB (status byte) query returns the current value
of the status byte of the instrument.
Description
The master summary status (MSS) bit 6 indicates
whether or not the device has at least one reason for
requesting service. When you read the status byte reg-
ister, the value returned is the total of the bit weights of
all of the bits set to one at the time you read the byte.
The following table shows each bit in the status byte
register and its bit weight. The *STB? query does not
affect the contents of the status byte register.
Query Response
<integer>
from 0 to 255.
Example
OUTPUT 720;”*STB?”
ENTER 720;Value
PRINT Value
*TRG
The *TRG (trigger) command is identical to the group
execute trigger (GET) message or RUN command.
Description
This command acquires data according to the current
settings. This command cannot be issued as a query. If a
measurement is already in progress, a trigger is ignored,
and an error is generated.
Bit Bit
Weight
Condition
7
128
Not Used
6
64
Master Summary Status (MSS)
5
32
Event Status Bit (ESB)
4
16
Message Available (MAV)
3
8
Not Used
2
4
Error queue status
1
2
Not Used
0
1
Not Used
Summary of Contents for 86121A
Page 2: ...User s Guide Agilent 86121A WDM Channel Analyzer ...
Page 7: ......
Page 10: ...Chapter 1 Getting Started ...
Page 21: ......
Page 24: ...Agilent 86121A Front and Rear Panels 2 3 Agilent 86121A Front and Rear Panels ...
Page 25: ...2 4 Agilent 86121A Front and Rear Panels ...
Page 27: ...2 6 WDM Channel Analyzer Display The S N with Averaging display ...
Page 32: ...The Softkeys 2 11 The Setup menu ...
Page 33: ...2 12 The Softkeys The Disk menu The Printer menu ...
Page 35: ...2 14 Changing the Printer Paper Changing the Printer Paper ...
Page 37: ......
Page 191: ......
Page 205: ...7 14 Regulatory Information Declaration of Conformity ...
Page 239: ......