5
-
12
Measurement Instructions
MEASure{:ARRay | [:SCALar]}:POWer? [<expected_value>[,<resolution>]]
Returns amplitude values.
Description
When used with a :SCALar command, a single value is
returned. The display is placed in the single-wavelength
mode, and the marker is placed on the signal having a
power level that is closest to the
<expected_value>
parameter.
When used with an :ARRay command, an array of
amplitudes is returned. The display is placed in the list-
by-power mode.
Returned values are in the current power units. Wave-
length units are not affected.
<expected_value>
Constants
Examples
:CONF:ARR:POW
:FETC:ARR:POW?
:READ:ARR:POW?
:MEAS:ARR:POW?
:CONF:SCAL:POW -10 dBm
:FETC:SCAL:POW? MAX
:READ:SCAL:POW? MIN
:MEAS:SCAL:POW? DEF
Used With
<expected_value>
<resolution>
SCALar
optional
ignored
ARRay
ignored
ignored
CONFigure command
When this function is used with the CONFigure command, the
query question mark character “?”
must
not
be included in the
string. However, the FETCh, READ, and MEASure command are
queries and require the question mark. Refer to the examples
for this command.
MAXimum
Displays the highest power signal
MINimum
Displays the lowest power signal
DEFault
Displays the signal at the current marker position
Summary of Contents for 86121A
Page 2: ...User s Guide Agilent 86121A WDM Channel Analyzer ...
Page 7: ......
Page 10: ...Chapter 1 Getting Started ...
Page 21: ......
Page 24: ...Agilent 86121A Front and Rear Panels 2 3 Agilent 86121A Front and Rear Panels ...
Page 25: ...2 4 Agilent 86121A Front and Rear Panels ...
Page 27: ...2 6 WDM Channel Analyzer Display The S N with Averaging display ...
Page 32: ...The Softkeys 2 11 The Setup menu ...
Page 33: ...2 12 The Softkeys The Disk menu The Printer menu ...
Page 35: ...2 14 Changing the Printer Paper Changing the Printer Paper ...
Page 37: ......
Page 191: ......
Page 205: ...7 14 Regulatory Information Declaration of Conformity ...
Page 239: ......