5
-
16
Measurement Instructions
<resolution>
Constants
Examples
:CONF:ARR:POW:WAV DEF, MAX
:FETC:ARR:POW:WAV? DEF, MIN
:READ:ARR:POW:WAV?
:MEAS:ARR:POW:WAV?
:CONF:SCAL:POW:WAV 1300NM, MAX
:FETC:SCAL:POW:WAV? 1300NM, MIN
:READ:SCAL:POW:WAV? 1300NM
:MEAS:SCAL:POW:WAV? 1300NM
Query Response
The following line is an example of a returned string
when :MEAS:SCAL:POW:WAV? MAX is sent:
+1.5529258E-006
If six laser lines are located and :MEAS:ARR:POW:WAV?
is sent, the following string could be returned. Notice
that the first returned number indicates the number of
laser-line values returned in the query.
6,+1.54488100E-006,+1.54648400E-006,+1.54809000E-
006,+1.54969900E-006,+1.55131100E-006,+1.55292600E-006
MAXimum
0.01 resolution (fast update)
MINimum
0.001 resolution (normal)
DEFault
Current resolution
Summary of Contents for 86121A
Page 2: ...User s Guide Agilent 86121A WDM Channel Analyzer ...
Page 7: ......
Page 10: ...Chapter 1 Getting Started ...
Page 21: ......
Page 24: ...Agilent 86121A Front and Rear Panels 2 3 Agilent 86121A Front and Rear Panels ...
Page 25: ...2 4 Agilent 86121A Front and Rear Panels ...
Page 27: ...2 6 WDM Channel Analyzer Display The S N with Averaging display ...
Page 32: ...The Softkeys 2 11 The Setup menu ...
Page 33: ...2 12 The Softkeys The Disk menu The Printer menu ...
Page 35: ...2 14 Changing the Printer Paper Changing the Printer Paper ...
Page 37: ......
Page 191: ......
Page 205: ...7 14 Regulatory Information Declaration of Conformity ...
Page 239: ......