SENSe Subsystem
delay, and the last value is sampled at +41.48 mm opti-
cal path delay. When
FAST
measurement update is
selected, the first data value is sampled at +20.74 mm
optical path delay, and the last value is sampled at
–20.74 mm optical path delay. The data value that cor-
responds to zero optical path delay is approximately,
but not exactly, located in the center of the time domain
data.
If your program is aborted or interrupted after sending
this query, the Agilent 86121A continues to process the
data but does not place it in the output buffer. Because
of the amount of data processed, the instrument
will
not respond
to any new commands in its input buffer
for 30 or 40 seconds
.
Query Response
The following string shows an example of the first few
measurements returned by this query.
+1.51000,+1.51000,+1.49000,+1.47949200E
+000,+1.50000,+1.53000,+1.50000,+1.47265
600E+000,+1.50000,+1.50000,+1.51000,+1.4
8000,+1.50000,+1.51000,+1.50000
,+1.48000,+1.50000
Notice that only values are returned to the computer.
There is no first value that indicates the number of val-
ues contained in the string as there is, for example, with
the FETCh, READ, and MEASure commands.
Summary of Contents for 86121A
Page 2: ...User s Guide Agilent 86121A WDM Channel Analyzer ...
Page 7: ......
Page 10: ...Chapter 1 Getting Started ...
Page 21: ......
Page 24: ...Agilent 86121A Front and Rear Panels 2 3 Agilent 86121A Front and Rear Panels ...
Page 25: ...2 4 Agilent 86121A Front and Rear Panels ...
Page 27: ...2 6 WDM Channel Analyzer Display The S N with Averaging display ...
Page 32: ...The Softkeys 2 11 The Setup menu ...
Page 33: ...2 12 The Softkeys The Disk menu The Printer menu ...
Page 35: ...2 14 Changing the Printer Paper Changing the Printer Paper ...
Page 37: ......
Page 191: ......
Page 205: ...7 14 Regulatory Information Declaration of Conformity ...
Page 239: ......