Test 5. Amplitude Accuracy and Linearity
6
-
11
Test 5. Amplitude Accuracy and Linearity
Equipment
Amplitude linearity is performed using the following
devices:
• 1550 nm DFB lasers (>0 dBm output power,
±
0.005 dB
stability for 10 minutes)
• Optical attenuator (Agilent 8156A or equivalent)
• Agilent 11896A polarization controller
• Optical power meter (Agilent 8153A/81532A or
equivalent)
Procedure
1
Turn on the laser and allow it to warm up.
2
Connect output of the laser to the input of the optical
attenuator.
3
Connect the output of the optical attenuator to the
optical input of the polarization controller.
4
Connect the optical output of the polarization controller
to the input of the optical power meter.
5
Configure the optical power meter for 1550 nm.
6
Adjust the optical attenuator so that the power meter
measures a power level of 0 dBm.
7
Enter the attenuator setting and the measured power
level in the first row of
. Be sure
to enter these values into the “Attenuator Settings” and
“Power Meter Readings” columns.
8
Enter the measured power on the following line:
Pwr: ____________
9
Change the attenuator in 1 dB steps as shown in
, and enter the attenuator settings and power
Polarization sensitivity
To ensure measurement accuracy, minimize the movement of
any fiber-optic cables during this procedure. Moving cables
causes polarization changes which affect amplitude measure-
ments.
Summary of Contents for 86121A
Page 2: ...User s Guide Agilent 86121A WDM Channel Analyzer ...
Page 7: ......
Page 10: ...Chapter 1 Getting Started ...
Page 21: ......
Page 24: ...Agilent 86121A Front and Rear Panels 2 3 Agilent 86121A Front and Rear Panels ...
Page 25: ...2 4 Agilent 86121A Front and Rear Panels ...
Page 27: ...2 6 WDM Channel Analyzer Display The S N with Averaging display ...
Page 32: ...The Softkeys 2 11 The Setup menu ...
Page 33: ...2 12 The Softkeys The Disk menu The Printer menu ...
Page 35: ...2 14 Changing the Printer Paper Changing the Printer Paper ...
Page 37: ......
Page 191: ......
Page 205: ...7 14 Regulatory Information Declaration of Conformity ...
Page 239: ......