Definition of Terms
7
-
3
Definition of Terms
Wavelength
Range
refers to the allowable wavelength range of the
optical input signal.
Absolute accuracy
indicates the maximum wavelength
error over the allowed environmental conditions. The
wavelength accuracy is based on fundamental physical
constants, which are absolute standards not requiring
traceability to artifacts kept at national standards labo-
ratories. Two He-Ne gas lasers are used. First, there is
an internal 632.991 nm-vacuum (473.6127 THz) refer-
ence laser
a
. To verify absolute wavelength accuracy, an
external laser is measured during manufacturing at
1523.488 nm, or 196.7804 THz
b
.
Differential Accuracy
indicates the maximum wave-
length error in measuring the wavelength difference
between two signals that are simultaneously present.
Minimum Resolvable Separation
indicates the mini-
mum wavelength separation of two laser lines required
to measure each wavelength simultaneously. Two laser
lines closer in wavelength than the minimum resolvable
separation are not resolved and one average wavelength
is displayed.
Display Resolution
indicates the minimum incremen-
tal change in displayed wavelength.
Amplitude
Calibration Accuracy
indicates the maximum power
calibration error at the specified wavelengths over the
allowed environmental conditions. The amplitude cali-
bration accuracy is traceable to a National Institute of
Standards and Technology (NIST) calibrated optical
power meter. NIST is the national standards laboratory
of the United States.
a. Obarski, G. E. 1990. “Wavelength Measurement System for Opti-
cal Fiber Communications”
NIST Technical Note 1336
(Febru-
ary): 18. Take the average of the two frequencies straddling gain
center.
b. D. A. Jennings, F. R. Peterson, and K. M. Evenson. 1979. “Fre-
quency measurement of the 260-THz (1.15 micron) He-Ne laser”
Optics Letters Vol. 4, No. 5
(May): 129-130.
Summary of Contents for 86121A
Page 2: ...User s Guide Agilent 86121A WDM Channel Analyzer ...
Page 7: ......
Page 10: ...Chapter 1 Getting Started ...
Page 21: ......
Page 24: ...Agilent 86121A Front and Rear Panels 2 3 Agilent 86121A Front and Rear Panels ...
Page 25: ...2 4 Agilent 86121A Front and Rear Panels ...
Page 27: ...2 6 WDM Channel Analyzer Display The S N with Averaging display ...
Page 32: ...The Softkeys 2 11 The Setup menu ...
Page 33: ...2 12 The Softkeys The Disk menu The Printer menu ...
Page 35: ...2 14 Changing the Printer Paper Changing the Printer Paper ...
Page 37: ......
Page 191: ......
Page 205: ...7 14 Regulatory Information Declaration of Conformity ...
Page 239: ......