Changing the Units and Measurement Rate
3
-
7
Changing the Units and Measurement Rate
Displayed units
To change the units of measure
1
Press
Setup
.
2
Press the
MORE
softkey.
3
Press the
WL UNIT
softkey.
4
Select one of the following units. Then, press
RETURN
to
complete your selection:
•
NM
for nanometers
•
THZ
for terahertz
Measurement rate
Under normal operation, the Agilent 86121A makes a
measurement and displays the results about once every
second. It is in this normal update mode that maximum
accuracy and wavelength resolution are achieved. How-
ever, should a faster update be desired, for example
when real-time feedback is required to tune a laser to
its designated channel, the 86121A can be set to update
approximately two times per second. This reduces both
wavelength resolution and accuracy but can be benefi-
cial in some applications.
The instrument resolution in
NORMAL
update mode is
7 GHz (0.06 nm at 1550 nm). This resolution is useful
when measuring closely spaced laser lines carrying data
at up to 5 Gb/s rates.
The instrument resolution in
FAST
update mode is
14 GHz (0.12 nm at 1550 nm). This resolution is useful
when measuring laser lines carrying data at up to
10 Gb/s rates.
Summary of Contents for 86121A
Page 2: ...User s Guide Agilent 86121A WDM Channel Analyzer ...
Page 7: ......
Page 10: ...Chapter 1 Getting Started ...
Page 21: ......
Page 24: ...Agilent 86121A Front and Rear Panels 2 3 Agilent 86121A Front and Rear Panels ...
Page 25: ...2 4 Agilent 86121A Front and Rear Panels ...
Page 27: ...2 6 WDM Channel Analyzer Display The S N with Averaging display ...
Page 32: ...The Softkeys 2 11 The Setup menu ...
Page 33: ...2 12 The Softkeys The Disk menu The Printer menu ...
Page 35: ...2 14 Changing the Printer Paper Changing the Printer Paper ...
Page 37: ......
Page 191: ......
Page 205: ...7 14 Regulatory Information Declaration of Conformity ...
Page 239: ......