Making Measurements
4
-
11
Always force the Agilent 86121A to wait for non-
sequential commands
The Agilent 86121A normally processes its remote pro-
gramming commands sequentially. The instrument
waits until the actions specified by a particular com-
mand are completely finished before reading and exe-
cuting the next command. However, there are a few
non-sequential commands where this is not true. Non-
sequential commands do
not
finish executing before
the next command is interpreted.
The following is a list of the non-sequential commands
of the Agilent 86121A:
:CALCulate1:TRANsform:FREQuency:POINTs
:CALCulate2:PEXCursion
:CALCulate2:PTHReshold
:CALCulate2:WLIMit:STARt:FREQuency
:CALCulate2:WLIMit:STARt:WAVelength
:CALCulate2:WLIMit:STOP:FREQuency
:CALCulate2:WLIMit:STOP:WAVelength
:SENSe:CORRection:ELEVation
:INITiate:CONTinuous
:INITiate[:IMMediate]
The benefit of non-sequential commands is that, in
some situations, they can reduce the overall execution
times of programs. For example, you can set the peak
excursion, peak threshold, and elevation and use a
*WAI command at the end to save time. However, non-
sequential commands can also be a source of annoying
errors. Always use the *OPC query or *WAI command
with the non-sequential commands to ensure that your
programs execute properly.
ARRay and the SCPI standard
According to the SCPI command reference, the ARRay com-
mand causes an instrument to take multiple measurements. (A
<size> parameter indicates the number of measurements to
take.) However, the Agilent 86121A’s ARRay command refers to
the measurements performed for one measurement sweep;
this results in an array of measured signals. Because the <size>
parameter does not apply, any <size> parameter sent will be
ignored by the instrument. No syntax error will be generated if
a <size> parameter is sent.
Summary of Contents for 86121A
Page 2: ...User s Guide Agilent 86121A WDM Channel Analyzer ...
Page 7: ......
Page 10: ...Chapter 1 Getting Started ...
Page 21: ......
Page 24: ...Agilent 86121A Front and Rear Panels 2 3 Agilent 86121A Front and Rear Panels ...
Page 25: ...2 4 Agilent 86121A Front and Rear Panels ...
Page 27: ...2 6 WDM Channel Analyzer Display The S N with Averaging display ...
Page 32: ...The Softkeys 2 11 The Setup menu ...
Page 33: ...2 12 The Softkeys The Disk menu The Printer menu ...
Page 35: ...2 14 Changing the Printer Paper Changing the Printer Paper ...
Page 37: ......
Page 191: ......
Page 205: ...7 14 Regulatory Information Declaration of Conformity ...
Page 239: ......