CALCulate1 Subsystem
5
-
17
CALCulate1 Subsystem
Use the CALCulate1 commands to query
uncorrected
frequency-spectrum data. In
NORMAL
measurement
update mode, 15,047 values are returned. If the
Agilent 86121A is set for
FAST
measurement update
mode (low resolution), 7,525 values are returned.
The commands in this subsystem have the following
command hierarchy:
:CALCulate1
:DATA?
:TRANsform
:FREQuency
:POINts
:CALCulate1:DATA?
Queries
uncorrected
frequency-spectrum data of the
input laser line.
Attribute Summary
Preset
State: not affected
SCPI Compliance: standard
Query Only
Description
The returned values are proportional to squared Watts
(linear) units. No amplitude or frequency correction is
applied to the values. To obtain the logarithmic (dB)
result, normalize the returned values by the largest
value, then take five times the logarithm of the normal-
ized values.
Be prepared to process a large amount of data when
this query is sent. The amount of data returned
depends on the measurement update state of the
instrument which can be set using the
:CALCulate1:TRANsform:FREQuency:POINts com-
mand or the resolution argument of an instrument func-
tion.
Refer to “Measurement Instructions” on page 5-11
.
When
NORMAL
measurement update is specified, over
250 kilobytes of data (15,047 values) can be returned to
the computer. When
FAST
measurement update is speci-
fied, over 125 kilobytes of data (7,525 values) can be
returned.
Summary of Contents for 86121A
Page 2: ...User s Guide Agilent 86121A WDM Channel Analyzer ...
Page 7: ......
Page 10: ...Chapter 1 Getting Started ...
Page 21: ......
Page 24: ...Agilent 86121A Front and Rear Panels 2 3 Agilent 86121A Front and Rear Panels ...
Page 25: ...2 4 Agilent 86121A Front and Rear Panels ...
Page 27: ...2 6 WDM Channel Analyzer Display The S N with Averaging display ...
Page 32: ...The Softkeys 2 11 The Setup menu ...
Page 33: ...2 12 The Softkeys The Disk menu The Printer menu ...
Page 35: ...2 14 Changing the Printer Paper Changing the Printer Paper ...
Page 37: ......
Page 191: ......
Page 205: ...7 14 Regulatory Information Declaration of Conformity ...
Page 239: ......