5
-
50
SENSe Subsystem
Instrument specifications apply when the device is set
to NARRow. Specifications do not apply in BROad
mode.
Query Response
The query form returns the previously selected device.
NARR
ow
:SENSe:CORRection:ELEVation{?| {<integer> | MINimum | MAXimum}}
Sets the elevation value used by the instrument to com-
pensate for air dispersion.
<integer>
is the altitude in meters.
Attribute Summary
Non-sequential command
Preset
State: unaffected by
*RST sets this value to the minimum
SCPI Compliance: instrument specific
Description
Changing the elevation value causes the current data to
be reprocessed.
Query Response
The query form returns the current elevation setting as
shown in the following example:
+1500
:SENSe:CORRection:OFFSet:MAGNitude{?| {<real> | MINimum | MAXimum}}
Enters an offset for amplitude values.
<real>
is the logarithmic units in dB.
Constant
Description
MINimum
0 m
MAXimum
5000 m
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that
this command has the time to complete before sending any
more commands to the instrument.
Agilent 86121A to wait for non-sequential commands” on
page 4-11
for more information.
Summary of Contents for 86121A
Page 2: ...User s Guide Agilent 86121A WDM Channel Analyzer ...
Page 7: ......
Page 10: ...Chapter 1 Getting Started ...
Page 21: ......
Page 24: ...Agilent 86121A Front and Rear Panels 2 3 Agilent 86121A Front and Rear Panels ...
Page 25: ...2 4 Agilent 86121A Front and Rear Panels ...
Page 27: ...2 6 WDM Channel Analyzer Display The S N with Averaging display ...
Page 32: ...The Softkeys 2 11 The Setup menu ...
Page 33: ...2 12 The Softkeys The Disk menu The Printer menu ...
Page 35: ...2 14 Changing the Printer Paper Changing the Printer Paper ...
Page 37: ......
Page 191: ......
Page 205: ...7 14 Regulatory Information Declaration of Conformity ...
Page 239: ......