Error Messages
8
-
7
–200
Execution error
–211
“Trigger in progress”
Caused by sending the *TRG command when the instrument is already
taking a measurement or when the instrument is in continuous
measurement mode.
–213
“Init in process”
Caused by sending an INIT:IMM, READ, or MEASure command while a
measurement is already in progress or while the instrument is in
continuous measurement mode.
–221
“Settings conflict”
Caused by trying to set the instrument to a state that is not allowed. For
example, turning on drift maximum and drift minimum state
simultaneously or turning on SNR state while drift or delta state is on.
–222
Data out of range
–223
Too much data
–224
Illegal parameter value
–230
“Data stale”
Caused by trying to query measurement data immediately after a *RST
command. For example, sending *RST; FETCh or sending *RST;
:CALC2:DATA? pow.
–232
“Data questionable”
Caused by sending a resolution value in one of the measurement
functions that is outside the range of the instrument.
–250
Mass storage error
–252
Media missing
–254
Media full
–255
Dir full
–257
File name error
–258
Media protected
–273
Illegal macro label
–310
System error
–321
Out of memory
–350
Too many errors
–400
Query error
Table 8-3. General SCPI Error Messages (2 of 3)
Error Number
Description
Summary of Contents for 86121A
Page 2: ...User s Guide Agilent 86121A WDM Channel Analyzer ...
Page 7: ......
Page 10: ...Chapter 1 Getting Started ...
Page 21: ......
Page 24: ...Agilent 86121A Front and Rear Panels 2 3 Agilent 86121A Front and Rear Panels ...
Page 25: ...2 4 Agilent 86121A Front and Rear Panels ...
Page 27: ...2 6 WDM Channel Analyzer Display The S N with Averaging display ...
Page 32: ...The Softkeys 2 11 The Setup menu ...
Page 33: ...2 12 The Softkeys The Disk menu The Printer menu ...
Page 35: ...2 14 Changing the Printer Paper Changing the Printer Paper ...
Page 37: ......
Page 191: ......
Page 205: ...7 14 Regulatory Information Declaration of Conformity ...
Page 239: ......