Setting Up the 86122A
Cleaning Connections for Accurate Measurements
46
Agilent 86122A
Multi-Wavelength Meter
, Second Edition
To clean a non-lensed connector
1
Apply pure isopropyl alcohol to a clean lint-free cotton swab or lens paper.
Cotton swabs can be used as long as no cotton fibers remain on the fiber end after
cleaning.
2
Clean the ferrules and other parts of the connector while avoiding the end of the
fiber.
3
Apply isopropyl alcohol to a new clean lint-free cotton swab or lens paper.
4
Clean the fiber end with the swab or lens paper.
Do not scrub during this initial cleaning because grit can be caught in the
swab and become a gouging element.
5
Immediately dry the fiber end with a clean, dry, lint-free cotton swab or lens
paper.
6
Blow across the connector end face from a distance of 6 to 8 inches using
filtered, dry, compressed air. Aim the compressed air at a shallow angle to the
fiber end face.
Nitrogen gas or compressed dust remover can also be used.
Table 7 Dust Caps Provided with Lightwave Instruments
Item
Agilent Part Number
Laser shutter cap
08145-64521
FC/PC dust cap
08154-44102
Biconic dust cap
08154-44105
DIN dust cap
5040-9364
HMS10/dust cap
5040-9361
ST dust cap
5040-9366
Do not use any type of foam swab to clean optical fiber ends. Foam
swabs can leave filmy deposits on fiber ends that can degrade
performance.
C A U T I O N
Do not shake, tip, or invert compressed air canisters, because this releases
particles in the can into the air. Refer to instructions provided on the
compressed air canister.
C A U T I O N
Summary of Contents for 86122A
Page 1: ...Agilent Technologies Agilent 86122A Multi Wavelength Meter User s Guide ...
Page 8: ...8 Agilent 86122A Multi Wavelength Meter Second Edition ...
Page 74: ...Using the 86122A Menus 74 Agilent 86122A Multi Wavelength Meter Second Edition The Help menu ...
Page 92: ...Using the 86122A Making Measurements 92 Agilent 86122A Multi Wavelength Meter Second Edition ...
Page 111: ......