5-20
Status Listings
OSA Warnings
5044
desc = “Please cycle power to synchronize system time”
help = “The system time clock has been set backwards. Due to internal system
requirements it is necessary to cycle power before continuing. Please turn power
off and back on again. This must be done to ensure proper system operation.”
5045
desc = “Sweep limit markers too close together”
help = “The sweep limit function is on and the line markers are too close together.
The OSA cannot limit the sweep range to such a small value. The current sweep
may be using a wider range than requested. Move the line markers farther apart or
turn off sweep limit to resolve this problem.”
5046
desc = “AutoAlign cannot find input signal at marker”
help = “The auto-align procedure could not find a usable input signal. If there is an
active marker auto-align will attempt to align at the marker wavelength without
searching for a signal peak. This error can also occur if the input signal amplitude is
very low or if the OSA has received a large mechanical shock or if the input signal is
removed after the auto-align procedure has begun. Be sure a signal is connected to
the optical input. If there is an active marker make sure it is placed on the input
signal or turn off all markers.”
5047
desc = “Signal disappeared during AutoAlign”
help = “The auto-align procedure did not complete. The input signal disappeared
during the auto-align procedure. Make sure the input signal is connected to the
optical input and try again.”
5048
desc = “Wrong trace X axis units for peak to center”
help = “The peak-to-center function was requested. This function will only execute
if the active trace X axis has units of meters (um nm pm). Change the active trace to
one with an X axis in meters or re-measure the active trace in a non-zero span.”
5049
desc = “Wrong marker X axis units for active trace”
help = “The active marker cannot be placed on the active trace because the
desired X axis units do not match the X axis units of the active trace. This is normally
caused by recalling a wavelength trace in zero span or a zero span trace in a
wavelength span. It can also be caused by loading a trace with conflicting units by
remote control.”
5050
desc = “Trace not displayed: wrong X axis units”
help = “View has been turned on for this trace but it cannot be displayed. The X
axis units in the trace do not match the X axis units being displayed. This is normally
caused by trying to view a wavelength trace in zero span or a zero span trace in a
wavelength span. It can also be caused by loading a trace with conflicting units by
remote control.”
Table 5-6. OSA Warnings (4 of 12)
Error Number
Error Description [description/explanation/examples]
Summary of Contents for 8614 B Series
Page 1: ...Agilent 8614xB Series Optical Spectrum Analyzer User s Guide ...
Page 12: ...1 4 Getting Started Product Overview Agilent 8614xB Front and Rear Panels ...
Page 13: ...1 5 Getting Started Product Overview ...
Page 15: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Page 26: ...1 18 Getting Started The Softkey Panels ...
Page 31: ...1 23 Getting Started The Softkey Panels The Systems Menus continued ...
Page 40: ...1 32 Getting Started Product Options and Accessories ...
Page 75: ...3 Function Reference ...
Page 186: ...4 16 Remote Front Panel Operation Remote Front Panel ...
Page 226: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...
Page 271: ......