5-32
Status Listings
Application-Specific Warnings
8022
desc = “A closing quote is missing.”
help = “The specification file contained a line with an opening quote but no closing
quote could be found.”
8023
desc = “A default path name was used.”
help = “The imported specification file was given a default path name. All
specification files must have at least one PATH statement before any of the
measurement statements. A default path name was used either because some
measurement statements were used before a PATH statement or there was no
PATH statement at all in the file.”
8024
desc = “A variable is expected.”
help = “The statement expected a variable as an input parameter. The line number
indicates which statement did not have the proper input parameter type.”
8025
desc = “The path name has already been used.”
help = “The application allows new names to be assigned to measurement paths.
Once assigned these names can no longer be used by other measurement paths.
Only the first thirty-two characters are significant. The line number indicates where
the problem was found. The word in the parentheses is the path name causing the
problem.”
8026
desc = “A path name was used where a variable is expected.”
help = “The input parameters for measurement statements can take variables as
inputs. These variables are the results of other measurement statements. A path
name is the name of a measurement path not the result of a measurement and
cannot be used as an input parameter. The input parameter was ignored for this
case. The line number indicates where the problem was found. The word in the
parentheses is the path name causing the problem.”
8027
desc = “A parameter value was outside the legal range.”
help = “A parameter value was used which was outside the allowable range. The
first number in the parentheses is the number of the parameter causing the
problem. The other two numbers are the minimum and maximum values allowed.
These values are in base units (e.g meters watts).”
8028
desc = “A minimum non-zero span is required.”
help = “The application cannot run in a zero span setting. The start and stop
wavelength must be separated by a minimum span. The line number indicates the
line in the specification file which had a start and stop wavelength pair that did not
meet the minimum span requirement.”
Table 5-7. Application-Specific Warnings (4 of 6)
Error Number
Error Description [description/explanation/examples]
Summary of Contents for 8614 B Series
Page 1: ...Agilent 8614xB Series Optical Spectrum Analyzer User s Guide ...
Page 12: ...1 4 Getting Started Product Overview Agilent 8614xB Front and Rear Panels ...
Page 13: ...1 5 Getting Started Product Overview ...
Page 15: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Page 26: ...1 18 Getting Started The Softkey Panels ...
Page 31: ...1 23 Getting Started The Softkey Panels The Systems Menus continued ...
Page 40: ...1 32 Getting Started Product Options and Accessories ...
Page 75: ...3 Function Reference ...
Page 186: ...4 16 Remote Front Panel Operation Remote Front Panel ...
Page 226: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...
Page 271: ......