Programmer’s Guide
10-19
Menu Map with SCPI Commands
Menu Map for 8712ET/ES and 8714ET/ES
Table 10-5
CAL
Functions, 8712ET and 8714ET (1 of 3)
1. Pressing this softkey begins a guided 1-port calibration procedure, using three
standards. When the calibration is complete, the command
*WAI;:SENS[1|2]
:CORR:COLL:SAVE;*WAI
is executed. The order of appearance and the content of
the softkey menus depend on the measurement channel 1 and measurement
channel 2.
KEYSTROKES
SCPI COMMAND
CAL,
(Reflection)
(hardkey entry)
Default 1-Port
SENS[1|2]:CORR:CSET DEF; *WAI
1-Port
SENS[1|2]:CORR:COLL:IST OFF;METH REFL3;
*WAI
Measure Standard
1
SENS[1|2]:CORR:COLL:IST OFF;METH REFL3;
*WAI
Measure Standard — Open
SENS[1|2]:CORR:COLL STAN1;*WAI
Measure Standard — Short
SENS[1|2]:CORR:COLL STAN2;*WAI
Measure Standard — Load
SENS[1|2]:CORR:COLLSTAN3;*WAI;
:SENS[1|2]:CORR:COLL:SAVE;*WAI
Normalize on OFF
TRAC CH[1|2]SMEM,CH[1|2]SDATA;:CALC[1|2]:
MATH (IMPL/CH[1|2]SMEM);:DISP:WIND[1|2]:
TRAC[1|2] ON;TRAC[1|2] OFF
Cal Check
(see
(menu selection only)
Summary of Contents for 8712ES
Page 11: ...1 1 1 Introduction to GPIB Programming ...
Page 27: ...2 1 2 Synchronizing the Analyzer and a Controller ...
Page 36: ...3 1 3 Passing Control ...
Page 39: ...4 1 4 Data Types and Encoding ...
Page 46: ...5 1 5 Using Status Registers ...
Page 71: ...6 1 6 Trace Data Transfers ...
Page 98: ...6 28 Programmer sGuide Trace Data Transfers Internal Measurement Arrays ...
Page 99: ...7 1 7 Using Graphics ...
Page 105: ...8 1 8 Front Panel Keycodes ...
Page 111: ...9 1 9 Introduction to SCPI ...
Page 129: ...10 1 10 Menu Map with SCPI Commands ...
Page 268: ...12 1 12 SCPI Conformance Information ...
Page 290: ...13 1 13 SCPI Error Messages ...