Chapter 1
1-11
Alphabetical Command Reference
ATT
ATT
Syntax
ATT<A|B|P1|P2><num>[DB];
or
ATT<A|B|P1|P2>?;
Description
Front Panel Equivalents
Command
Description
Range
Query Response
ATTA
1
1. This command only applies to 8720E series analyzers with Option 085.
Selects the amount of attenuation at
attenuator A.
0–55 dB
<num><
L
F
>
ATTB
Selects the amount of attenuation at
attenuator B.
0–55 dB
<num><
L
F
>
ATTP1
2
2. This command only applies to 8753ES Option 011 analyzers.
Selects the amount of attenuation at
PORT 1.
0–70 dB
<num><
L
F
>
ATTP2
Selects the amount of attenuation at
PORT 2.
0–70 dB
<num><
L
F
>
Command
Hardkey
Softkey
ATTA
1
1. This command only applies to 8720E series analyzers with Option 085.
ATTB
ATTP1
2
2. This command only applies to 8753ES Option 011 analyzers.
ATTP2
Power
ATTENUATOR A
Power
ATTENUATOR B
Power
ATTENUATOR PORT 1
Power
ATTENUATOR PORT 2
Summary of Contents for 8719ES