1-96
Chapter 1
Alphabetical Command Reference
FWD
FWD
Syntax
FWD<I|M|T>;
Description
These commands are OPC-compatible if there is only one standard in the class. If there is
just one standard, that standard is measured automatically. If there is more than one
standard in the class, the standard being used must be selected with the
STAN
command.
Front Panel Equivalents
Command
Description
Range
Query Response
FWDI
Selects the forward isolation calibration
class during a 2-port calibration sequence.
N/A
N/A
FWDM
Selects the forward match calibration
class during a 2-port calibration sequence.
N/A
N/A
FWDT
Selects the forward transmission
calibration class during a 2-port
calibration sequence.
N/A
N/A
Command
Hardkey
Softkey
FWDI
FWDM
FWDT
Cal
FWD ISOL’N
Cal
FWD MATCH
Cal
FWD TRANS