Chapter 4
4-5
Reading Analyzer Data
Marker Data
Table 4-1 Units as a Function of Display Format
Display
Format
Marker
Mode
OUTPMARK
OUTPFORM
Marker Readout
1
1. The marker readout values are the marker values displayed in the upper right-hand corner of the display.
They also correspond to the value and auxiliary value associated with the fixed marker.
Value 1
Value 2
Value 1
Value 2
Value
Aux
Value
LOG MAG
dB
N/S
2
2. Value 2 is not significant in this format, though it is included in data transfers. See also
Transfer Commands” on page 5-5
.
dB
N/S
dB
N/S
PHASE
degrees
N/S
degrees
N/S
degrees
N/S
DELAY
seconds
N/S
seconds
N/S
seconds
N/S
SMITH
CHART
LIN MKR
lin mag
degrees
real
imag
lin mag
degrees
LOG MKR
dB
degrees
real
imag
dB
degrees
Re/Im
real
imag
real
imag
real
imag
R + jX
real
ohms
imag
ohms
real
imag
real
ohms
imag
ohms
G + jB
real
Siemens
imag
Siemens
real
imag
real
Siemens
imag
Siemens
POLAR
LIN MKR
lin mag
degrees
real
imag
lin mag
degrees
LOG MKR
dB
degrees
real
imag
dB
degrees
Re/Im
real
imag
real
imag
real
imag
LIN MAG
lin mag
N/S
lin mag
N/S
lin mag
N/S
SWR
SWR
N/S
SWR
N/S
SWR
N/S
REAL
real
N/S
real
N/S
real
N/S
IMAGINARY
imag
N/S
imag
N/S
imag
N/S
Summary of Contents for 8719ES
Page 15: ...1 1 1 Alphabetical Command Reference ...
Page 293: ...2 1 2 Introduction to Instrument Control ...
Page 310: ...3 1 3 GPIB Programming ...
Page 334: ...4 1 4 Reading Analyzer Data ...
Page 343: ...5 1 5 Data Processing Chain ...
Page 350: ...6 1 6 Error Reporting ...
Page 364: ...7 1 7 Programming Examples ...
Page 502: ...A 1 A Preset Conditions ...
Page 517: ...B 1 B Command Listings ...