Chapter 3
3-11
Adjustments and Correction Constants
RF Output Power Correction Constants (Test 47)
RF Output Power Correction Constants (Test 47)
Required Equipment and Tools
Analyzer warm-up time: 30 minutes.
This procedure adjusts several correction constants that can improve the output power
level accuracy of the internal source. They are related to the power level, power slope,
power slope offset, and the ALC roll-off factors among others.
1. If you just completed
“Sampler Magnitude and Phase Correction Constants (Test 53),”
start at step 8 in this procedure.
2. Press
.
3. Press
. The default power meter
address is 13. Refer to the power meter manual as required to observe or change its
GPIB address.
4. Press
to toggle between the 438A/437 and 436A power meters.
Choose the appropriate model number. (Use the 438A/437 selection if the power meter
is an E4419B or E4418B.)
NOTE
If you are using the 438A power meter, connect the 8482A power sensor to
channel A, and the 8481A power sensor to channel B.
Description
HP/Agilent Part or Model Number
Power Meter
436A/437B/438A or E4418B/4419B
GPIB Cable
10833A
Antistatic Wrist Strap
9300-1367
Antistatic Wrist Strap Cord
9300-0980
Static-control Table Mat and Ground Wire
9300-0797
Additional Equipment for 50
Ω
Analyzers
Power Sensor
8482A
Power Sensor (for Option 006 analyzers)
8481A
Adapter APC-7 to Type-N (f)
(for 8753ES only)
11524A
Additional Equipment for 75
Ω
Analyzers
Power Sensor
8483A Option H03
Preset
Local
SYSTEM CONTROLLER
Local
SET ADDRESSES ADDRESS: P MTR/GPIB
POWER MTR:438A/437
Summary of Contents for 8753ES
Page 14: ...Contents xiv Contents ...
Page 15: ...1 1 1 Service Equipment and Analyzer Options ...
Page 26: ...1 12 Chapter1 Service Equipment and Analyzer Options Service and Support Options ...
Page 27: ...2 1 2 System Verification and Performance Tests ...
Page 203: ...3 1 3 Adjustments and Correction Constants ...
Page 262: ...3 60 Chapter3 Adjustments and Correction Constants Sequences for Mechanical Adjustments ...
Page 263: ...4 1 4 Start Troubleshooting Here ...
Page 297: ...5 1 5 Power Supply Troubleshooting ...
Page 317: ......
Page 318: ...6 1 6 Digital Control Troubleshooting ...
Page 337: ...6 20 Chapter6 Digital Control Troubleshooting GPIB Failures ...
Page 338: ...7 1 7 Source Troubleshooting ...
Page 369: ...7 32 Chapter7 Source Troubleshooting Source Group Troubleshooting Appendix ...
Page 370: ...8 1 8 Receiver Troubleshooting ...
Page 381: ...8 12 Chapter8 Receiver Troubleshooting Troubleshooting When One or More Inputs Look Good ...
Page 382: ...9 1 9 Accessories Troubleshooting ...
Page 389: ...9 8 Chapter9 Accessories Troubleshooting Inspect the Error Terms ...
Page 390: ...10 1 10 Service Key Menus and Error Messages ...
Page 439: ...10 50 Chapter10 Service Key Menus and Error Messages Error Messages ...
Page 440: ...11 1 11 Error Terms ...
Page 451: ...11 12 Chapter11 Error Terms Error Correction ...
Page 452: ...12 1 12 Theory of Operation ...
Page 482: ...13 1 13 Replaceable Parts ...
Page 487: ...13 6 Chapter13 Replaceable Parts Ordering Information Figure 13 1 Module Exchange Procedure ...
Page 500: ...Chapter 13 13 19 Replaceable Parts Replaceable Part Listings Figure 13 7 8753ET Cables Top ...
Page 502: ...Chapter 13 13 21 Replaceable Parts Replaceable Part Listings Figure 13 8 8753ES Cables Top ...
Page 512: ...Chapter 13 13 31 Replaceable Parts Replaceable Part Listings Figure 13 13 8753ES Cables Front ...
Page 544: ...14 1 14 Assembly Replacement and Post Repair Procedures ...
Page 550: ...Chapter 14 14 7 Assembly Replacement and Post Repair Procedures Covers Figure 14 2 Covers ...
Page 597: ...14 54 Chapter14 Assembly Replacement and Post Repair Procedures Post Repair Procedures ...