10-44
Chapter 10
Service Key Menus and Error Messages
Error Messages
Error Messages
This section contains an alphabetical list of the error messages that pertain to servicing
the analyzer. The information in the list includes explanations of the displayed messages
and suggestion to help solve the problem.
NOTE
The error messages that pertain to measurement applications are included in
the your analyzer’s reference guide.
BATTERY FAILED. STATE MEMORY CLEARED
Error Number 183
The battery protection of the non-volatile SRAM memory has failed. The SRAM
memory has been cleared. Refer to
Chapter 14 , “Assembly Replacement and
for battery replacement instructions. See the chapter
titled “Preset State and Memory Allocation” in the analyzer’s reference guide for more
information about the SRAM memory.
BATTERY LOW! STORE SAVE REGS TO DISK
Error Number 184
The battery protection of the non-volatile SRAM memory is in danger of failing. If this
occurs, all of the instrument state registers stored in SRAM memory will be lost. Save
these states to a disk and refer to
Chapter 14 , “Assembly Replacement and
for battery replacement instructions. See the chapter
titled “Preset State and Memory Allocation” in the analyzer’s reference guide for more
information about the SRAM memory.
CALIBRATION ABORTED
Error Number 74
You have changed the active channel during a calibration so the calibration in
progress was terminated. Make sure the appropriate channel is active and restart the
calibration.
CALIBRATION REQUIRED
Error Number 63
A calibration set could not be found that matched the current stimulus state or
measurement parameter. You will have to perform a new calibration.
CORRECTION CONSTANTS NOT STORED
Error Number 3
A store operation to the EEPROM was not successful. You must change the position of
the switch on the A9 CPU assembly. Refer to
“A9 Switch Positions” on page 3-6
.
CORRECTION TURNED OFF
Error Number 66
Critical parameters in your current instrument state do not match the parameters for
the calibration set, therefore correction has been turned off. The critical instrument
state parameters are sweep type, start frequency, frequency span, and number of
points.
Summary of Contents for 8753ES
Page 14: ...Contents xiv Contents ...
Page 15: ...1 1 1 Service Equipment and Analyzer Options ...
Page 26: ...1 12 Chapter1 Service Equipment and Analyzer Options Service and Support Options ...
Page 27: ...2 1 2 System Verification and Performance Tests ...
Page 203: ...3 1 3 Adjustments and Correction Constants ...
Page 262: ...3 60 Chapter3 Adjustments and Correction Constants Sequences for Mechanical Adjustments ...
Page 263: ...4 1 4 Start Troubleshooting Here ...
Page 297: ...5 1 5 Power Supply Troubleshooting ...
Page 317: ......
Page 318: ...6 1 6 Digital Control Troubleshooting ...
Page 337: ...6 20 Chapter6 Digital Control Troubleshooting GPIB Failures ...
Page 338: ...7 1 7 Source Troubleshooting ...
Page 369: ...7 32 Chapter7 Source Troubleshooting Source Group Troubleshooting Appendix ...
Page 370: ...8 1 8 Receiver Troubleshooting ...
Page 381: ...8 12 Chapter8 Receiver Troubleshooting Troubleshooting When One or More Inputs Look Good ...
Page 382: ...9 1 9 Accessories Troubleshooting ...
Page 389: ...9 8 Chapter9 Accessories Troubleshooting Inspect the Error Terms ...
Page 390: ...10 1 10 Service Key Menus and Error Messages ...
Page 439: ...10 50 Chapter10 Service Key Menus and Error Messages Error Messages ...
Page 440: ...11 1 11 Error Terms ...
Page 451: ...11 12 Chapter11 Error Terms Error Correction ...
Page 452: ...12 1 12 Theory of Operation ...
Page 482: ...13 1 13 Replaceable Parts ...
Page 487: ...13 6 Chapter13 Replaceable Parts Ordering Information Figure 13 1 Module Exchange Procedure ...
Page 500: ...Chapter 13 13 19 Replaceable Parts Replaceable Part Listings Figure 13 7 8753ET Cables Top ...
Page 502: ...Chapter 13 13 21 Replaceable Parts Replaceable Part Listings Figure 13 8 8753ES Cables Top ...
Page 512: ...Chapter 13 13 31 Replaceable Parts Replaceable Part Listings Figure 13 13 8753ES Cables Front ...
Page 544: ...14 1 14 Assembly Replacement and Post Repair Procedures ...
Page 550: ...Chapter 14 14 7 Assembly Replacement and Post Repair Procedures Covers Figure 14 2 Covers ...
Page 597: ...14 54 Chapter14 Assembly Replacement and Post Repair Procedures Post Repair Procedures ...