230
Chapter 16
Diagnostics
Troubleshooting a LA
1
interface problem
Troubleshooting a LA
1
interface problem
If the 89600 software cannot find your measurement hardware, the problem may
be in the LAN interface.
To verify the correct ADC.
1. Click
Utilities > Hardware.
2. If your measurement hardware is not included in the list for ADC1, continue
with “To troubleshoot the LAN interface.”
3. If your measurement hardware is included in the list for ADC1, click to
highlight it.
4. Make sure that the
Default Hardware
and
Simulate Hardware
checkboxes are
cleared.
5. Click
OK.
6. If the analyzer still does not use your hardware, continue with “To troubleshoot
the LAN interface.”
To troubleshoot the LA
1
interface:
1. Close all 89600 applications. To verify that all applications are closed, run Task
Manager.
a. Right click an empty space in the task bar or press Alt-Ctrl-Delete.
b. Click
Task Manager … > Applications.
c. If there are any
AgtVSAVector
or
AgtVSAScalar
applications listed,
highlight them and click
End Task.
d. Click the
Processes
tab.
e. If there are any
VSAVEC*.EXE
or
VSASCA*.EXE
processes in the list,
highlight them and click
End Task.
2. Confirm that your PC can communicate with the instrument.
a. Open a Command Prompt window: for Windows XP, click
Start > All
Programs > Accessories > Command Prompt.
b. Type
ping hostname
and press Enter,
where hostname is the IP address or
hostname for the instrument. (For instructions on finding the IP address or
hostname, see
“Instrument IP address for a crossover cable configuration”
.)
If you are successful, continue with
If you are not successful,
continue with the next sub-step.
c. Make sure that the PC is on and the LAN cable is connected to the PC and
Summary of Contents for 89604
Page 2: ...Agilent Technologies Agilent 89600 Series Installation and VXI Service Guide ...
Page 14: ...13 1 Installation Overview ...
Page 23: ...22 Chapter1 Installation Overview 89600 Series Analyzer Installation and Setup ...
Page 24: ...23 2 Installing 89600 Software ...
Page 32: ...31 3 Setting Up VXI Hardware ...
Page 77: ...76 Chapter3 Setting Up VXI Hardware Connecting EMC grounding cable ...
Page 78: ...77 4 Installing IEEE 1394 Interfaces in Computers ...
Page 86: ...85 5 Installing GPIB LA1 and USB Interface Cards ...
Page 93: ...92 Chapter5 Installing GPIB LAN and USB Interface Cards Configuring a LA1 to GPIB gateway ...
Page 94: ...93 6 Configuring IO Interfaces ...
Page 112: ...111 7 Configuring Infiniium Windows XP Scopes ...
Page 126: ...125 8 Installing 89601 Software in Agilent Infiniium XP Scopes ...
Page 132: ...131 9 Installing Agilent 89600 Software in Agilent X Series Signal Analyzers ...
Page 138: ...137 10 Linking Logic Analyzers to the Agilent 89600 VSA ...
Page 152: ...Chapter 10 151 Linking Logic Analyzers to the Agilent 89600 VSA Logic Analyzer Link Examples ...
Page 156: ...Chapter 10 155 Linking Logic Analyzers to the Agilent 89600 VSA Logic Analyzer Link Examples ...
Page 161: ...160 Chapter10 Linking Logic Analyzers to the Agilent 89600 VSA Logic Analyzer Link Examples ...
Page 162: ...161 11 Linking a Synthetic Instrument SI Spectrum Analyzer to the Agilent 89600 VSA ...
Page 170: ...169 12 Connecting Acqiris Waveform Digitizers to the Agilent 89600 VSA ...
Page 188: ...187 13 Installing a 1ode Locked License ...
Page 198: ...197 14 Setting Up a Floating License Server ...
Page 216: ...215 15 Setting Up a Floating License Client ...
Page 222: ...221 16 Diagnostics ...
Page 254: ...253 17 Circuit Descriptions ...
Page 277: ...276 Chapter18 Replacing VXI Hardware Storing VXI modules ...
Page 281: ...280 Index Index ...
Page 282: ...281 ...