Chapter 3
59
Setting Up VXI Hardware
Connecting front panel cables
Cable connections, two-channel 89611
NOTE
If you use external trigger, you must connect the trigger signal to the channel 1
ADC (E1439) trigger input.
The interconnections for the 10 MHz reference signals can be different, but you
must ensure that all the hardware is referenced to one signal.
1
2
3
4
5
3
5
4
6
External down converter 1
70 MHz
IF Out
Ext Ref In
Input
External down converter 2
70 MHz
IF Out
Ext Ref In
Input
89605B
89605B
E1439
E1439C
7
8
89611A
Ext Ref Out
10
9
Description
Connection
Part number
1
orange SMB-SMB
cable (100 mm)
E1439 #1
Clock
to E1439 #2
Clock
03585-61603
2
orange SMB-SMB
cable (100 mm)
E1439 #1
Sync
to E1439 #2
Sync
03585-61603
3
50 ohm SMB term (Qty
4)
Unused E1438
Clock
&
Sync
connectors
1250-0676
4
black BNC-BNC cable
(215 mm)
89605
To ADC Analog In
to E1439
Analog In
8121-0132
5
black BNC-BNC cable
(215 mm)
from 89605
10 MHz Ref Out
to E1439
Ext Clock/Ref
8121-0132
6
black BNC-BNC cable
(215 mm)
from 89605 1
10 MHz Ref Out
to89605 2
Ext Ref In
8121-0132
7
varies
external down converter
70 MHz IF Out
to
89605 Ch 1
Input
user supplied
8
varies
input to down converter
user supplied
9
varies
from 89605 #2
10 MHz Ref Out
to external down converter
#1
Ext Ref In
user supplied
Summary of Contents for 89604
Page 2: ...Agilent Technologies Agilent 89600 Series Installation and VXI Service Guide ...
Page 14: ...13 1 Installation Overview ...
Page 23: ...22 Chapter1 Installation Overview 89600 Series Analyzer Installation and Setup ...
Page 24: ...23 2 Installing 89600 Software ...
Page 32: ...31 3 Setting Up VXI Hardware ...
Page 77: ...76 Chapter3 Setting Up VXI Hardware Connecting EMC grounding cable ...
Page 78: ...77 4 Installing IEEE 1394 Interfaces in Computers ...
Page 86: ...85 5 Installing GPIB LA1 and USB Interface Cards ...
Page 93: ...92 Chapter5 Installing GPIB LAN and USB Interface Cards Configuring a LA1 to GPIB gateway ...
Page 94: ...93 6 Configuring IO Interfaces ...
Page 112: ...111 7 Configuring Infiniium Windows XP Scopes ...
Page 126: ...125 8 Installing 89601 Software in Agilent Infiniium XP Scopes ...
Page 132: ...131 9 Installing Agilent 89600 Software in Agilent X Series Signal Analyzers ...
Page 138: ...137 10 Linking Logic Analyzers to the Agilent 89600 VSA ...
Page 152: ...Chapter 10 151 Linking Logic Analyzers to the Agilent 89600 VSA Logic Analyzer Link Examples ...
Page 156: ...Chapter 10 155 Linking Logic Analyzers to the Agilent 89600 VSA Logic Analyzer Link Examples ...
Page 161: ...160 Chapter10 Linking Logic Analyzers to the Agilent 89600 VSA Logic Analyzer Link Examples ...
Page 162: ...161 11 Linking a Synthetic Instrument SI Spectrum Analyzer to the Agilent 89600 VSA ...
Page 170: ...169 12 Connecting Acqiris Waveform Digitizers to the Agilent 89600 VSA ...
Page 188: ...187 13 Installing a 1ode Locked License ...
Page 198: ...197 14 Setting Up a Floating License Server ...
Page 216: ...215 15 Setting Up a Floating License Client ...
Page 222: ...221 16 Diagnostics ...
Page 254: ...253 17 Circuit Descriptions ...
Page 277: ...276 Chapter18 Replacing VXI Hardware Storing VXI modules ...
Page 281: ...280 Index Index ...
Page 282: ...281 ...