70
Chapter 3
Setting Up VXI Hardware
Connecting front panel cables
Cable connections, 89641 with second RF channel
NOTE
If you use external trigger, you must connect the trigger signal to the channel 1
ADC (E1439) trigger input.
It is important that you use the semi-rigid SMA cables shipped with the analyzer.
Using other cables could degrade measurement accuracy.
1
2
4
5
6
7
8
9
3
4
10
11
8
5
3
6
7
12
13
1
2
Description
Connection
Part number
1
semi-rigid SMA cable
E2731
FE RF Out
to E2731
2.7GHz RF In
E2731-61601
2
semi-rigid SMA cable
89605
To RF In
to E2731
6 GHz In
89605-61694
3
N-to-BNC adapter
89605
Ch 1 Input
(user signal)
1250-0780
4
50 ohm SMB term (Qty
4)
Unused E1439
Clock
&
Sync
connectors
1250-0676
5
black BNC-BNC cable
(215 mm)
89605
To ADC Analog In
to E1439
Analog In
8121-0132
6
semi-rigid SMA cable
E2731
70 MHz IF Out
to 89605
From 70
MHz IF Out
89605-61691
7
semi-rigid SMA cable
89605
10 MHz Ext Ref Out
to E2731
10 MHz
Ext Ref
for 89605B, 89605-61693
for 89605A, 89605-61691
8
black BNC-BNC cable
(215 mm)
89605
10 MHz Ref Out
to E1439
Ext
Clock/Ref
8121-0132
9
black BNC-BNC cable
(215 mm)
89605 #1
10 MHz Ref Out
to 89605 #2
Ext
Ref In
8121-0132
10
orange SMB-SMB cable
(100 mm)
E1438 #1
Clock
to E1438 #2
Clock
03585-61603
11
orange SMB-SMB cable
(100 mm)
E1438 #1
Sync
to E1438 #2
Sync
03585-61603
12
semi-rigid SMA cable
E2731 #1 2nd LO In/Out to E2731 #2 2nd
LO In/Out
89605-61696
Summary of Contents for 89604
Page 2: ...Agilent Technologies Agilent 89600 Series Installation and VXI Service Guide ...
Page 14: ...13 1 Installation Overview ...
Page 23: ...22 Chapter1 Installation Overview 89600 Series Analyzer Installation and Setup ...
Page 24: ...23 2 Installing 89600 Software ...
Page 32: ...31 3 Setting Up VXI Hardware ...
Page 77: ...76 Chapter3 Setting Up VXI Hardware Connecting EMC grounding cable ...
Page 78: ...77 4 Installing IEEE 1394 Interfaces in Computers ...
Page 86: ...85 5 Installing GPIB LA1 and USB Interface Cards ...
Page 93: ...92 Chapter5 Installing GPIB LAN and USB Interface Cards Configuring a LA1 to GPIB gateway ...
Page 94: ...93 6 Configuring IO Interfaces ...
Page 112: ...111 7 Configuring Infiniium Windows XP Scopes ...
Page 126: ...125 8 Installing 89601 Software in Agilent Infiniium XP Scopes ...
Page 132: ...131 9 Installing Agilent 89600 Software in Agilent X Series Signal Analyzers ...
Page 138: ...137 10 Linking Logic Analyzers to the Agilent 89600 VSA ...
Page 152: ...Chapter 10 151 Linking Logic Analyzers to the Agilent 89600 VSA Logic Analyzer Link Examples ...
Page 156: ...Chapter 10 155 Linking Logic Analyzers to the Agilent 89600 VSA Logic Analyzer Link Examples ...
Page 161: ...160 Chapter10 Linking Logic Analyzers to the Agilent 89600 VSA Logic Analyzer Link Examples ...
Page 162: ...161 11 Linking a Synthetic Instrument SI Spectrum Analyzer to the Agilent 89600 VSA ...
Page 170: ...169 12 Connecting Acqiris Waveform Digitizers to the Agilent 89600 VSA ...
Page 188: ...187 13 Installing a 1ode Locked License ...
Page 198: ...197 14 Setting Up a Floating License Server ...
Page 216: ...215 15 Setting Up a Floating License Client ...
Page 222: ...221 16 Diagnostics ...
Page 254: ...253 17 Circuit Descriptions ...
Page 277: ...276 Chapter18 Replacing VXI Hardware Storing VXI modules ...
Page 281: ...280 Index Index ...
Page 282: ...281 ...