152
Chapter 5 Remote Interface Reference
Status System Commands
5
The Status Byte Register
The Status Byte Register reports conditions from the other register
groups. Data in the 3499A/B/C output buffer is reported on the MAV bit
(bit 4). If an event is cleared in one of the other registers, it is also cleared
in the Status Byte register. The
*CLS
command clears the Status Byte
register.
*STB?
This query reads the status byte. The returned value is a
decimal-weighted summary of the bits in the register.
*SRE <value>
*SRE?
These commands work with the enable register. The enable register is a
mask used to determine which bits in the event register can be
summarized to the RQS bit (bit 6) of the Status Byte Register.
value can range from 1 to 128. The query form of this command returns a
decimal-weighted value indicating the currently set bits in this register.
Summary of Contents for Agilent 3499A
Page 13: ...1 1 Quick Start ...
Page 27: ...2 2 Front Panel Operation ...
Page 55: ...3 3 System Overview ...
Page 77: ...4 4 Features and Functions ...
Page 113: ...5 5 Remote Interface Reference ...
Page 164: ...164 5 ...
Page 165: ...6 6 Error Messages ...
Page 173: ...7 7 Plug in Modules ...
Page 256: ...256 Chapter 7 Plug in Modules 44475A Breadboard Module 7 ...
Page 261: ...261 Chapter 7 Plug in Modules 44476B Microwave Switch Module 4 7 The 44476B is shown below ...
Page 286: ...286 Chapter 7 Plug in Modules Terminals and Connections Information 7 ...
Page 288: ...288 7 ...
Page 289: ...8 8 Application Programs ...
Page 299: ...9 9 Specifications ...
Page 343: ...343 Chapter 9 Specifications 44475A Breadboard Module 4 9 ...