231
Chapter 7 Plug-in Modules
44470A 10-Channel MUX Module
4
7
Custom Signal Conditioning
The 44470A circuit board has provision that allow you to install simple
attenuators or filter networks. Three circuit pads in each relay path are
provided that allow you to install components in the signal Hi path, Lo
path, or as a shunt from Hi to Lo. The figure below shows the 44470A
circuit board locations where components can be installed.
Common Bu
CH09
CH08
CH06
CH04
CH03
CH01
CH00
CH05
Common Bu
CH02
H
L
H
L
H
L
H
L
H
L
H
L
H
L
L
H
L
H
L
H
L
H CH07
H CH00
L
H CH09
L
H CH08
L
H CH06
L
H CH04
L
H CH03
L
H CH01
L
L
H CH05
L
H CH02
L
H CH07
44470ARelay MUX
Terminal
Block
LOW SERIES
HIGH SERIES
SHUNT
Summary of Contents for Agilent 3499A
Page 13: ...1 1 Quick Start ...
Page 27: ...2 2 Front Panel Operation ...
Page 55: ...3 3 System Overview ...
Page 77: ...4 4 Features and Functions ...
Page 113: ...5 5 Remote Interface Reference ...
Page 164: ...164 5 ...
Page 165: ...6 6 Error Messages ...
Page 173: ...7 7 Plug in Modules ...
Page 256: ...256 Chapter 7 Plug in Modules 44475A Breadboard Module 7 ...
Page 261: ...261 Chapter 7 Plug in Modules 44476B Microwave Switch Module 4 7 The 44476B is shown below ...
Page 286: ...286 Chapter 7 Plug in Modules Terminals and Connections Information 7 ...
Page 288: ...288 7 ...
Page 289: ...8 8 Application Programs ...
Page 299: ...9 9 Specifications ...
Page 343: ...343 Chapter 9 Specifications 44475A Breadboard Module 4 9 ...