128
Chapter 5 Remote Interface Reference
Switch Commands
5
[ROUTe:]
CLOSe:STATe?
This query returns a comma-delineated list of the closed channel relays
for all relays on all installed plug-in modules. The returned values will
be in the form snn, where s is the slot number and nn is the channel
number. The returned values indicate the programmed state of the relay.
This query does not physically monitor a relay and cannot report a failed
relay.
[ROUTe:]
OPEN <channel_list>|ALL
OPEN? <channel_list>
This command opens all the switching channels specified in the
channel_list
.
The channel_list has the form @snn, where s is the slot
number and nn is the channel number. See the
CLOSe
command
description for a description of channel_list.
The
OPEN?
query returns a “1” for open channels and a “0” for a closed
channel. If you specify a list of channels, the query returns a comma
separated list of 0’s and 1’s, in the same order as the channel_list. The
returned values indicate the programmed state of the relay. This query
does not physically monitor a relay and cannot report a failed relay.
Summary of Contents for Agilent 3499A
Page 13: ...1 1 Quick Start ...
Page 27: ...2 2 Front Panel Operation ...
Page 55: ...3 3 System Overview ...
Page 77: ...4 4 Features and Functions ...
Page 113: ...5 5 Remote Interface Reference ...
Page 164: ...164 5 ...
Page 165: ...6 6 Error Messages ...
Page 173: ...7 7 Plug in Modules ...
Page 256: ...256 Chapter 7 Plug in Modules 44475A Breadboard Module 7 ...
Page 261: ...261 Chapter 7 Plug in Modules 44476B Microwave Switch Module 4 7 The 44476B is shown below ...
Page 286: ...286 Chapter 7 Plug in Modules Terminals and Connections Information 7 ...
Page 288: ...288 7 ...
Page 289: ...8 8 Application Programs ...
Page 299: ...9 9 Specifications ...
Page 343: ...343 Chapter 9 Specifications 44475A Breadboard Module 4 9 ...