277
Chapter 7 Plug-in Modules
Terminals and Connections Information
4
7
N229X Screw Terminal Blocks
This terminal block is compatible with the N2290A, N2291A, N2292A,
N2293A, and N2295A plug-in modules. Refer to each plug-in module
section for detailed wiring information. The figure below shows the basic
steps to wire and assemble this terminal block.
STEP 1. Remove cover
A. Release screws on top of the cover;
B. Press tab forward and release.
STEP 2. Attach wires
A. The wire gauge: 18-26 AWG;
B. Insert wire into terminal connectors;
C. Tighten screws on the connectors.
STEP 3. Tighten the wires
A. Insert the wires into the nearest cable
boot, move the cable boot to the
Cable slot;
B. Tighten the cable boot.
STEP 4. Replace cover
A. Hook the top cover tabs onto the fixture;
B. Press down and tighten screws.
Cable Boot