1.1.1 Check Water flow and Gas flow errors
(1) Select MassHunter Instrument Control Window
→
Main menu
→
“View”
→
“Instrument
Status” to display “Instrument Status” Pane
If Main menu is not displayed, press F10 key.
(2) Click the dropdown below and check the error number for errors that are generated at the
date/time of the attempt to ignite plasma.
(3) Are error numbers 1440, 1445, or 1446 included in the latest errors?
Note
Summary of Contents for Agilent 7700 Series ICP-MS
Page 15: ...2 4 6 Check torch box 6 Check torch stand Torch stand with clean contact Previous Next Good ...
Page 19: ...2 7 1 Fix torch box issues 1 Replace or maintain the parts 2 Did this fix the problem Yes No ...
Page 23: ...2 11 1 Check cooling water 1 Replace water with pure water 2 Did this fix the problem Yes No ...
Page 25: ...3 2 1 Check if photo sensor cable is damaged 1 Is the Cable for photo sensor damaged Yes No ...
Page 34: ...4 8 1 Check usage of User Tune 1 Is User Tune function in use Yes No ...
Page 40: ...6 2 1 Drain spray chamber 1 Pull out drain tube and drain liquid from the spray chamber Next ...
Page 50: ...9 1 End of procedure Close this window Return to top ...