5-9
Programming Commands
Common Commands
Table 5-4. Conditions Set by *RST Reset
Item
Setting
Display mode
single wavelength
Wavelength range limiting
on
Start wavelength
1200 nm
Stop wavelength
1650 nm
Graphical display
off
Measurement acquisition
single
Wavelength calibration
vacuum
Elevation correction value
0 meters
Wavelength units
nm
Amplitude units
dBm
Power offset
0 dB
Peak threshold
10 dB
Peak excursion
15 dB
Measurement speed
normal
Number of uncorrected data points
34123
Delta Measurements:
∆
power
off
∆
wavelength
off
∆
wavelength and power
off
reference signal position
700 nm
Drift measurements
off
Coherence length measurements
off