7-8
Specifications and Regulatory Information
Specifications
Selectivity
Two lines input separated by
≥
100 GHz
(characteristic)
25 dB (characteristic)
Two lines input separated by
≥
30 GHz
(characteristic)
10 dB (characteristic)
Input Power
Maximum displayed level (sum of all lines)
+10 dBm
Maximum safe input level (sum of all lines)
+18 dBm
Maximum Number of Laser Lines Input
100
Input Return Loss
With flat contacting connectors
35 dB
With angled contacting connectors (Option 022)
50 dB
Measurement Cycle Time
Normal update mode
(characteristic)
1.0 s (1 measurement-per-second)
Fast update mode
(characteristic)
0.33 s (3 measurements-per-second)
Measurement Applications
Signal-to-Noise, channel spacing
≥
200 GHz (characteristic)
>35 dB (0.1 nm noise bandwidth), lines above –25
dBm
Signal-to-noise with Averaging, modulated lasers, 100 averages, channel
spacing
≥
200 GHz (characteristic)
>35 dB (0.1 nm noise bandwidth) lines above -25
dBm
Coherence Length
(characteristic)
Fabry-Perot lasers, 1 to 200 mm coherence length,
accuracy
±
5%, 0.75s cycle time
Summary of Contents for Agilent 86120B
Page 1: ...Agilent 86120B Multi Wavelength Meter User s Guide ...
Page 11: ...Contents 3 Contents Power Cords 8 16 Agilent Technologies Service Offices 8 18 ...
Page 12: ......
Page 38: ......
Page 68: ......
Page 100: ...4 18 Programming Monitoring the Instrument ...
Page 132: ......
Page 230: ......
Page 248: ...8 6 Reference Menu Maps Delta On Menu Delta Off Menu ...
Page 250: ...8 8 Reference Menu Maps System Setup Menu ...
Page 268: ......