2-14
Using the Multi-Wavelength Meter
Changing the Units and Measurement Rate
4
Press
WL
and select one of the following units. Then, press
RETURN
to complete
your selection:
•
NM
for nanometers
•
THZ
for terahertz
•
CM
–1
for wave number
5
Press
POWER
and select one of the following units:
•
DBM
for decibels relative to a milliwatt
•
MW
for milliwatts
•
UW
for microwatts
Measurement rate
Under normal operation, the Agilent 86120B makes a measurement and dis-
plays the results about once every second. It is in this normal update mode
that maximum accuracy and wavelength resolution are achieved. However,
should a faster update be desired, for example when real-time feedback is
required to tune a laser to its designated channel, the Agilent 86120B can be
set to update approximately three times per second. This reduces both wave-
length resolution and accuracy but can be beneficial in some applications.
When
FAST
update is selected, one less digit of resolution is displayed. Also, if
multiple wavelengths are present, these individual responses, with the
reduced resolution, may no longer be recognized.
To change the measurement speed
1
Press the
Setup
key.
2
Press the
MORE
softkey.
3
Press the
UPDATE
softkey.
4
Select either
NORMAL
or
FAST
.
Summary of Contents for Agilent 86120B
Page 1: ...Agilent 86120B Multi Wavelength Meter User s Guide ...
Page 11: ...Contents 3 Contents Power Cords 8 16 Agilent Technologies Service Offices 8 18 ...
Page 12: ......
Page 38: ......
Page 68: ......
Page 100: ...4 18 Programming Monitoring the Instrument ...
Page 132: ......
Page 230: ......
Page 248: ...8 6 Reference Menu Maps Delta On Menu Delta Off Menu ...
Page 250: ...8 8 Reference Menu Maps System Setup Menu ...
Page 268: ......