5-36
Programming Commands
CALCulate2 Subsystem
Description
When the state is on, the CALC2:DATA? POW query returns the total power
and the CALC2:DATA? WAV, FREQ, or WNUM query returns the power-
weighted average wavelength, frequency, or wave number values.
Turning power-weighted average mode on while making delta, coherence
length, or signal-to-noise measurements results in a “–221 Settings conflict”
error.
WLIMit[:STATe]
Limits input wavelength range of the Agilent 86120B.
Syntax
:CALCulate2:WLIMit[:STATe]{?| {ON | OFF | 1 | 0}}
Attribute
Summary
Non-sequential command
Preset
State: on
*RST State: on
SCPI Compliance: instrument specific
Description
When this function is on, the Agilent 86120B has an input range from the
WLIMit STARt to the WLIMit STOP. When this function is off, the instrument
displays peaks over the full wavelength range. If you want to measure signals
over a narrower wavelength range, set this function on to avoid identifying
spurious second harmonic peaks.
Whenever the Agilent 86120B receives this command, it reprocesses the data
and performs a new peak search.
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command has the
time to complete before sending any more commands to the instrument.
“Always force the Agilent 86120B to wait for non-sequential commands” on page 4-12
for more information.
Summary of Contents for Agilent 86120B
Page 1: ...Agilent 86120B Multi Wavelength Meter User s Guide ...
Page 11: ...Contents 3 Contents Power Cords 8 16 Agilent Technologies Service Offices 8 18 ...
Page 12: ......
Page 38: ......
Page 68: ......
Page 100: ...4 18 Programming Monitoring the Instrument ...
Page 132: ......
Page 230: ......
Page 248: ...8 6 Reference Menu Maps Delta On Menu Delta Off Menu ...
Page 250: ...8 8 Reference Menu Maps System Setup Menu ...
Page 268: ......