5-41
Programming Commands
CALCulate2 Subsystem
WLIMit:STOP[:WAVelength]
Sets the stopping wavelength for the wavelength limit range.
Syntax
:CALCulate2:WLIMit:STOP[WAVelength] {?|{<real> | MINimum | MAXimum }}
<
real
> is a wavelength value that is within the following limits:
Constant
Description
MINimum start
wavelength
limit
MAXimum 1650.0
nm
Attribute
Summary
Non-sequential command
Preset
State: 1650 nm
*RST State: 1650 nm
SCPI Compliance: instrument specific
Description
This command sets the stopping range for the wavelength limit. The default
units for the <
real
> parameter are meters. The stop wavelength value must be
greater than or equal to the start wavelength value or the stop wavelength will
be clipped to the start wavelength and a “Data out of range” error will be gen-
erated. Setting the start wavelength is equivalent to setting the start fre-
quency/wavenumber because of the inverse relationship of frequency to
wavelength.
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command has the
time to complete before sending any more commands to the instrument.
“Always force the Agilent 86120B to wait for non-sequential commands” on page 4-12
for more information.
Summary of Contents for Agilent 86120B
Page 1: ...Agilent 86120B Multi Wavelength Meter User s Guide ...
Page 11: ...Contents 3 Contents Power Cords 8 16 Agilent Technologies Service Offices 8 18 ...
Page 12: ......
Page 38: ......
Page 68: ......
Page 100: ...4 18 Programming Monitoring the Instrument ...
Page 132: ......
Page 230: ......
Page 248: ...8 6 Reference Menu Maps Delta On Menu Delta Off Menu ...
Page 250: ...8 8 Reference Menu Maps System Setup Menu ...
Page 268: ......