5-72
Programming Commands
SENSe Subsystem
CORRection:MEDium
Sets the Agilent 86120B to return wavelength readings in a vacuum or stan-
dard air.
Syntax
:SENSe:CORRection:MEDium{?| {AIR | VACuum}}
Argument
Description
AIR
Selects wavelength values in standard air.
VACuum
Selects wavelength values in a vacuum.
Attribute
Summary
Preset
State: VAC
*RST State: VAC
SCPI Compliance: instrument specific
Description
Standard air is defined to have the following characteristics:
Barometric pressure: 760 torr
Temperature: 15
°
C
Relative humidity: 0%
CORRection:OFFSet[:MAGNitude]
Enters an offset for amplitude values.
Syntax
:SENSe:CORRection:OFFSet:MAGNitude{?| {<real> | MINimum | MAXimum}}
<real>
is the logarithmic units in dB.
Constant
Description
MINimum
−
40.0 dB
MAXimum 40.0
dB
Attribute
Summary
Preset
State: 0.0
*RST State: 0.0
SCPI Compliance: standard