5-85
Programming Commands
TRIGger Subsystem
INITiate:CONTinuous
Selects single or continuous measurement acquisition.
Syntax
:INITiate:CONTinuous{?| {ON | OFF | 1 | 0}}
Attribute
Summary
Non-sequential command
Preset
State: on
*RST State: off
SCPI Compliance: standard
Description
When on is specified, the instrument continuously measures the input spec-
trum.
INITiate[:IMMediate]
Initiates a new measurement sequence.
Syntax
:INITiate:IMMediate
Attribute
Summary
Non-sequential command
Preset
State: none
SCPI Compliance: standard
Command Only
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command has the
time to complete before sending any more commands to the instrument.
“Always force the Agilent 86120B to wait for non-sequential commands” on page 4-12
for more information.