6-4
Performance Tests
Test 2. Sensitivity
Test 2. Sensitivity
Description
Sensitivity is verified using the following devices:
• Optical power meter
• Optical attenuator
• 1310 nm and 1550 nm lasers (>0 dBm output power)
C A U T I O N
Do not 18 dBm source power. The Agilent 86120B’s input circuitry can
be damaged when total input power exceeds 18 dBm.
Procedure
Perform the following procedure first using the 1310 nm laser and then repeat
the steps using the 1550 nm laser.
1
Connect the laser’s output to the optical attenuator’s input.
2
Connect the optical attenuator’s output to the optical power meter.
3
Adjust the attenuator for a reading of 0 dBm on the power meter. Record the
attenuator’s setting.
Attenuation at 0 dBm: ____________
4
Adjust the attenuator for a reading of –35 dBm on the power meter. Record the
attenuator’s setting.
Attenuation at –35 dBm: ____________
5
Disconnect the fiber-optic cable at the power meter’s input and connect the
cable to the Agilent 86120B being tested.
6
Reset the optical attenuator to the setting recorded in
Step 3
.
7
Read the power and wavelength measured on the Agilent 86120B, and
compared them to the specifications listed in
Chapter 7, “Specifications and
.
8
Reset the optical attenuator to the setting recorded in
Step 4
.
9
Read the power and wavelength measured on the Agilent 86120B, and
compared them to the specifications listed in
Chapter 7, “Specifications and
.
Summary of Contents for Agilent 86120B
Page 1: ...Agilent 86120B Multi Wavelength Meter User s Guide ...
Page 11: ...Contents 3 Contents Power Cords 8 16 Agilent Technologies Service Offices 8 18 ...
Page 12: ......
Page 38: ......
Page 68: ......
Page 100: ...4 18 Programming Monitoring the Instrument ...
Page 132: ......
Page 230: ......
Page 248: ...8 6 Reference Menu Maps Delta On Menu Delta Off Menu ...
Page 250: ...8 8 Reference Menu Maps System Setup Menu ...
Page 268: ......