Multimeter Command Reference
FUNCtion
[SENSe:]FUNCtion
“<function>”
selects the measurement function. You can
select the functions shown in the following table.
Parameters
Comments
*RST
Condition:
SENS:VOLT:DC
Example
Change Measurement Function
CONF:VOLT
Function: DC voltage.
FUNC “FRES”
Set function to 4-wire resistance.
READ?
Place multimeter in wait-for-trigger state and
make measurement; send reading to output
buffer.
enter statement
Enter reading into computer.
FUNCtion?
[SENSe:]FUNCtion?
returns one of the following quoted strings to the output
buffer:
“
CURR:AC
”
“
CURR
”
“
FREQ
”
“
FRES
”
“
PER
”
“
RES
”
“
VOLT:AC
”
“
VOLT
”
“
VOLT:RAT
”
Example
Query the Measurement Function
FUNC “FRES”
Function: 4-wire ohms.
FUNC?
Query multimeter to return selected function.
enter statement
Enter quoted string into computer.
Parameter Name
Parameter Type
Range of Values
Default Units
<
function
>
discrete
:CURRent:AC|
:CURRent[:DC]|
:FREQuency|
:FRESistance|
:PERiod|
:RESistance|
:VOLTage:AC|
:VOLTage[:DC]|
:VOLTage[:DC]:RATio
none