Multimeter Command Reference
143
RESistance:RESolution
[SENSe:]RESistance:RESolution <
resolution
>
selects the resolution for 2-wire
resistance measurements.
Parameters
Comments
•
MINimum
selects the best resolution (the smallest value) for the selected range.
MAXimum
selects the worst resolution (the largest value) for the selected range.
•
You must select a range using
RESistance:RANGe
before specifying
resolution. Also, only specify a resolution when making measurements on a
fixed range. Otherwise, the resolution will change to correspond with the range
selected during autoranging.
•
If autoranging is required, set the resolution using the
MIN
or
MAX
parameters.
•
If necessary to achieve the specified resolution, the multimeter will increase
the integration time as needed. This command overrides the results of
previously executed
RESistance:NPLC
or
RESistance:APERture
command
(the last command executed has priority).
•
The
RES:RESolution
command overrides the resolution setting from a
previous
CONFigure:RESistanc
e command.
•
Related Commands:
CONFigur
e,
RESistance:NPLC
•
*RST
Condition:
Based on the
*RST
values for the
RESistance:NPLC
command.
Example
Change the Resolution
CONF:RES 1560,MAX
Function: 2-wire ohms; range selected: 10k
;
MAX resolution: 1
.
RES:RES 10E-03
Set resolution to 10m
.
READ?
Place multimeter in wait-for-trigger state and make
measurements; send readings to output buffer.
enter statement
Enter readings into computer.
RESistance:RESolution?
[SENSe:]RESistance:RESolution? [MIN|MAX]
returns one of the following
numbers to the output buffer.
•
The present resolution selected if
MIN
or
MAX
are not specified. Only the
resolution values available on ranges set by the
RANGe
command are
returned.
•
The resolution with the smallest value (i.e., the best resolution) for the selected
range if
MIN
is specified.
•
The resolution with the largest value (i.e., the worst resolution) for the selected
range if
MAX
is specified.
Parameter Name
Parameter Type
Range of Values
Default Units
<
resolution
>
numeric
resolution
|MIN|MAX
ohms
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...