148
Multimeter Command Reference
VOLTage[:DC]:NPLC
[SENSe:]VOLTage[:DC]:NPLC <
number
>
sets the integration time in power line
cycles (PLCs). The NPLC is set to a value from the range of values that can
accommodate the
<
number
>
specified. 11 sets NPLC to 100.
Parameters
Comments
•
MIN
selects 0.02 PLCs.
MAX
selects 100 PLCs. Setting the integration time in
PLCs also sets the aperture time and the resolution. For example, 10 PLCs sets
an aperture time of 167ms (60Hz line frequency) or 200ms (50Hz). The
corresponding resolution depends on the function and range you select.
•
The
VOLT:DC:NPLC
command overrides the results of previously executed
VOLTage:DC:RESolution
commands (the last command executed has
priority).
•
The greater the number of PLCs, the greater the normal mode rejection (and
the lower the reading rate).
•
Only the 1 PLC, 10 PLC and 100 PLC settings provide normal mode rejection
of 50Hz or 60Hz power line related noise.
•
*RST
Condition:
10 PLC
Example
Set the Integration Time in PLCs
VOLT:DC:NPLC 10
Integration time is 10 PLCs.
VOLTage[:DC]:NPLC?
[SENSe:]VOLTage[:DC]:NPLC? [MIN|MAX]
returns one of the following numbers
to the output buffer:
•
The present integration time in PLCs if
MIN
or
MAX
is not specified.
•
The minimum integration time available (0.02) if
MIN
is specified.
•
The maximum integration time available (100) if
MAX
is specified.
Example
Query the Integration Time
VOLT:DC:NPLC 100
Integration time is 100 PLCs.
VOLT:DC:NPLC?
Query multimeter to return integration time.
enter statement
Enter value into computer.
Parameter Name
Parameter Type
Range of Values
Default Units
<
number
>
numeric
0.02|0.2|1|10|100|MIN|MAX
PLCs
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...