Agilent E1412A Multimeter Error Messages
185
Appendix B
Agilent E1412A Multimeter Error Messages
The following sections describe the types of errors the E1412A reports;
Execution Errors, Self-Test Errors and Calibration Errors. The error code is
given (e.g., -101) followed by the associated error message and a description
of what the error message means.
Execution Errors
-101
Invalid character
An invalid character was found in the command string. You may have
inserted a character such as #, $, or % in the command header or within a
parameter. Example:
CONF:VOLT#DC
-102
Syntax error
Invalid syntax was found in the command string. You may have inserted a
blank space before or after a colon in the command header, or before a
comma. Example:
SAMP:COUN ,1
-103
Invalid separator
An invalid separator was found in the command string. You may have used
a comma instead of a colon, semicolon, or blank space – or you may have
used a blank space instead of a comma. Example:
TRIG:COUN,1
or
CONF:FREQ 1000 0.1
-104
Data type error
The wrong parameter type was found in the command string. You may have
specified a number where a string was expected, or vice versa.
Example:
TRIG:COUN '150'
or
TRIG:COUN A
-105
GET not allowed
A Group Execute Trigger (GET) is not allowed within a command string.
-108
Parameter not allowed
More parameters were received than expected for the command. You may
have entered an extra parameter, or you added a parameter to a command
that does not accept a parameter. Example:
READ? 10
-109
Missing parameter
Fewer parameters were received than expected for the command. You
omitted one or more parameters that are required for this command.
Ex
ample: SAMP:COUN
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...