186
Agilent E1412A Multimeter Error Messages
-112
Program mnemonic too long
A command header was received which contained more than the maximum
12 characters allowed. Example:
CONFIGURATION:VOLT:DC
A
<
code
>
string contains more than the maximum 12 characters allowed in
the
CALibration:SECure:CODE
command.
-113
Undefined header
A command was received that is not valid for this multimeter. You may have
misspelled the command or it may not be a valid command. If you are using
the short form of the command, remember that it may contain up to four
letters. Example:
TRIGG:COUN 3
-121
Invalid character in number
An invalid character was found in the number specified for a parameter
value. Example:
STAT:QUES:ENAB #B01010102
-123
Numeric overflow
A numeric parameter was found whose exponent was larger than 32,000.
Example:
TRIG:COUN 1E34000
-124
Too many digits
A numeric parameter was found whose mantissa contained more than
255 digits, excluding leading zeros.
-128
Numeric data not allowed
A numeric parameter was found but a character string was expected. Check
the list of parameters to verify you have used a correct parameter type.
Example:
TRIG:SOUR 1
-131
Invalid suffix
A suffix was incorrectly specified for a numeric parameter. You may have
misspelled the suffix. Example:
TRIG:DEL 0.5 SECS
-138
Suffix not allowed
A suffix was received following a numeric parameter which does not accept
a suffix. Example:
SAMP:COUN 1 SEC
(
SEC
is not a valid suffix).
-148
Character data not allowed
A character string was received but a numeric parameter was expected.
Check the list of parameters to verify that you have used a valid parameter
type. Example:
CAL:LFR XYZ
-151
Invalid string data
An invalid character string was received. Check to see if you have enclosed
the character string in single or double quotes. Example:
CAL:STR 'NEXT
CAL DUE 10/4/1996
(the ending quote is missing).
-158
String data not allowed
A character string was received but is not allowed for the command. Check
the list of parameters to verify that you have used a valid parameter type.
Example:
CALC:STAT 'ON'
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...