Agilent E1412A Multimeter Error Messages
187
-160 to -168
Block data errors
The multimeter does not accept block data.
-170 to -178
Expression errors
The multimeter does not accept mathematical expressions.
-211
Trigger ignored
A Group Execute Trigger (GET) or
*TRG
was received but the trigger was
ignored. Make sure the multimeter is in the “wait-for-trigger” state before
issuing a trigger, and make sure the correct trigger source is selected.
-213
Init ignored
An
INITiate
command was received but could not be executed because a
measurement was already in progress. Send a device clear to halt a
measurement in progress and place the multimeter in the “idle” state.
-214
Trigger deadlock
A trigger deadlock occurs when the trigger source is
BUS
and a
READ?
command is received.
-221
Settings conflict
This error can be generated in one of the following situations:
You sent a
CONFigure
or
MEASure
command with autorange enabled and
with a fixed resolution. Example:
CONF:VOLT:DC DEF,0.1
You turned math on (
CALC:STAT ON
) and then changed to a math operation
that was not valid with the present measurement function. For example, dB
measurements are not allowed with 2-wire ohms. The math state is turned
off as a result of this condition.
-222
Data out of range
A numeric parameter value is outside the valid range for the command.
Example:
TRIG:COUN -3
-223
Too much data
A character string was received but could not be executed because the string
length was more than 12 characters. This error can be generated by the
CALibration:STRing
and
DISPlay:TEXT
commands.
-224
Illegal parameter value
A discrete parameter was received which was not a valid choice for the
command. You may have used an invalid parameter choice.
Examples:
CALC:FUNC SCALE
(
SCALE
is not a valid choice) or
SAMP:COUN ON
(
ON
is not a valid choice).
-230
Data stale
A
FETCh?
command was received but internal reading memory was empty.
The reading retrieved may be invalid or settings have changed since the data
was taken.
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...