188
Agilent E1412A Multimeter Error Messages
-330
Self-test failed
The multimeter's complete self-test failed from the remote interface (
*TST?
command). In addition to this error, more specific self-test errors are also
reported.
See also “Self-Test Errors,” following this section.
-350
Too many errors
The error queue is full because more than 20 errors have occurred. No
additional errors are stored until you remove errors from the queue. The
error queue is cleared when power has been off, or after a
*CLS
(clear status)
command has been executed.
-410
Query
INTERRUPTED
A command was received which sends data to the output buffer, but the
output buffer contained data from a previous command (the previous data is
not overwritten). The output buffer is cleared when power has been off, or
after a
*RST
(reset) command has been executed.
-420
Query
UNTERMINATED
The multimeter was addressed to talk (i.e., to send data over the interface)
but a command has not been received which sends data to the output buffer.
For example, you may have executed a
CONFigure
command (which does
not generate data) and then attempted an
ENTER
statement to read data from
the remote interface.
-430
Query
DEADLOCKED
A command was received which generates too much data to fit in the output
buffer and the input buffer is also full. Command execution continues but all
data is lost.
-440
Query
UNTERMINATED
after indefinite response
The
*IDN?
command must be the last query command within a command
string. Example:
*IDN?;:SYST:VERS?
501
Isolator UART framing error
502
Isolator UART overrun error
511
Unexpected reset occurred
The outguard circuit recognized the inguard circuit reset (probably due to an
abnormal input condition). This error causes the instrument to go to the
power-on setting and the previous setting is lost.
521
Input buffer overflow
522
Output buffer overflow
531
Insufficient memory
There is not enough memory to store the requested number of readings in
internal memory using the
INITiate
command. The product of the sample
count (
SAMPle:COUNt
) and the trigger count (
TRIGger:COUNt
) must not
exceed 512 readings.
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...