Multimeter Command Reference
INPut
The
INPut
command enables or disables the automatic input impedance mode for DC
voltage measurements.
Subsystem Syntax
INPut
:IMPedance:AUTO OFF|ON
:IMPedance:AUTO?
:IMPedance:AUTO
INPut:IMPedance:AUTO <
mode
>
enables or disables the automatic input
impedance mode for DC voltage measurements. When disabled (
AUTO OFF
), the
multimeter maintains its input impedance of 10M
for all DC voltage ranges. This
is useful to prevent a change in input impedance, caused by changing ranges, from
affecting the measurements.
Parameters
Example
Enable Automatic Input Impedance (use >10G
for 100mV, 1V and 10V ranges)
INP:IMP:AUTO ON
Enable automatic input impedance.
Comments
•
You can substitute decimal values for the
OFF
(“
0
”) and
ON
(“
1
”) parameters.
•
*RST
Conditions:
INP:IMP:AUTO OFF
:IMPedance:AUTO?
INPut:IMPedance:AUTO?
returns a number to show whether the automatic input
impedance mode is enabled or disabled: “
1
” =
ON
, “
0
” =
OFF
. The number is sent
to the output buffer.
Example
Query the Input Impedance Mode
INP:IMP:AUTO ON
Enable automatic input impedance.
INP:IMP:AUTO?
Query multimeter to return input impedance
mode (“1”).
enter statement
Enter value into computer.
Parameter Name
Parameter Type
Range of Values
Default Units
<
mode
>
boolean
OFF|0|ON|1
None
mode
(Impedance)
AUTO OFF
(10M
)
AUTO ON (>10G
)
Range for
Impedance
all ranges
100mV, 1V and 10V
(other ranges are at 10M
)
Summary of Contents for Agilent E1412A
Page 2: ......
Page 10: ...8 Contents Notes...
Page 14: ...12 Notes...
Page 15: ...13 Notes...
Page 16: ...14 Notes...
Page 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Page 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Page 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...